Melting Point Apparatus: How It Works and How to Choose One

Melting Point Apparatus: How It Works and How to Choose One

A melting point test is one of the quickest ways to check the identity and purity of a solid. A pure crystalline substance melts sharply at a characteristic temperature; impurities usually lower the melting point and make it melt over a wider range. That is why melting point apparatus is found in almost every pharmaceutical QC lab, pharmacy college and chemistry teaching lab.

How the capillary method works

  1. A small amount of dry, finely powdered sample is packed into a thin glass capillary tube, closed at one end, to a height of a few millimetres.
  2. The capillary is placed in a heated block or oil bath next to a thermometer or temperature sensor.
  3. The temperature is raised quickly to a few degrees below the expected melting point, then slowly, typically about 1°C per minute.
  4. You watch the sample through a magnifier and record the temperature when it starts to melt and when it is completely liquid. That is the melting range.

Tips for accurate results

  • Dry and powder the sample well; moisture and coarse crystals widen the melting range.
  • Pack evenly and keep the sample height consistent between tests.
  • Heat slowly near the melting point. Heating too fast is the most common cause of high readings.
  • Check calibration regularly with reference standards of known melting point.
  • Follow your pharmacopoeia or in-house method for ramp rate and reporting.

Basic or digital apparatus?

Our PT-122 melting point apparatus is a simple, economical instrument for melting points up to 300°C using capillary tubes. It suits teaching labs and routine checks. Capillaries are available in 75 mm (PT-122/10) and 100 mm (PT-122/15) lengths, in packs of 100.

Our PT-125 digital melting point apparatus is the better choice for QC labs. It has:

  • a range from 2°C above room temperature to 300°C, readable to 1°C;
  • a digital temperature indicator with PT-100 sensor;
  • a built-in magnetic stirrer that keeps the oil bath at an even temperature;
  • an electronic heating-rate controller with HIGH/LOW range;
  • a magnifier and glare-free illuminated background for clear viewing;
  • an SS 304 body, supplied with 175 ml of silicone oil and a box of capillaries.

Quick comparison

PT-122 PT-125 digital
Range Up to 300°C Room temperature +2°C to 300°C
Temperature reading Thermometer Digital, PT-100 sensor, 1°C
Heating-rate control Basic Electronic, HIGH/LOW
Bath stirring No Built-in magnetic stirrer
Best for Teaching labs, routine checks Pharmaceutical QC, R&D, contract testing

Frequently asked questions

Why does an impure sample melt at a lower temperature?

Impurities disrupt the crystal structure, so the solid starts melting earlier and melts over a wider range. A sharp melting range close to the reference value is a good sign of purity.

What heating rate should I use?

Heat quickly to a few degrees below the expected melting point, then at about 1°C per minute, or as your pharmacopoeia or method specifies.

What is the maximum temperature of your melting point apparatus?

Both the PT-122 and the PT-125 measure melting points up to 300°C.

Which capillaries do I need?

Glass capillaries closed at one end. We supply 75 mm and 100 mm capillaries in packs of 100.

How do I get a quotation?

Use “Make an enquiry” on the product page, WhatsApp 8199900060 or email popular1942@gmail.com.